| METROLAB - 950 |
|
Elisa strip reader. 50 methods in memory. Curve up to 10 standards. Cut-off. Universal holder strip.
- Dynamic range: -0.3000 to 3.000 DO
- Spectral range: 400 to 700 nm
- Bandwidth: 10nm with filters provided by factory
- Stray light: <0.2 %
- Photometric linearity: better than +/- 0.005 A @ 0.4 A
- Detector: UV enhanced silicon photodiode
- Display: alfanumeric LCD, 16 characters
- Keyboard: soft touch. 16 multifunction keys
- Buil-in programs: Diagnostic and parameter verification. Optimum lamp adjustment. Absorbance (direct photometric reading). Factor and standard. Calibration curve with 10 standards and curve fit up to 21 functions
- Sample holder: universal 8 or 12 wells strips
- Output: serial RS232C for PC ( cable X950CA02). Serial for Epson compatible printer (cable X950CA01)
- Power: includes external power supply with input 110/220 Volts +/- 10 % - 50/60 Hz.
- Dimensions: 45 mm height; 150 mm width; 180 mm depth
- Weight: 550 grams
|
|
|
|
METROLAB - 330
UV-Vis spectrophotometer Range 330-1000 nm. Bandwidth 8 nm. T%. Absorbance. Concentration.
- Optical system: Single beam
- Wavelength range: 330-1000 nm
- Bandwidth: 8 nm
- Accuracy: +/- 2 nm
- Precision: +/-1 nm
- Monochromator: Czerny Turner holographic grating with protective coating
- Detector: Wide range silicon photodiode
- Stray light: <0.2 % @ 340 nm
- Photometric presentation
- Transmittance: 0.0 to 199.9 % of T
- Absorbance: 0.000 to 1.999 A
- Concentration: 0.000 to 1999 C
- Photometric linearity: +/- 1 % T
- Photometric accuracy: +/- 0.002 @ 0.4 A
- Stability: better than 0.002 A / hour after 15 min. in 0 A
- Automatic filters selection: yes
- Automatic Zero adjustment: yes
- Sample tubes: diameter 12 mm x 50mm
- Cuvettes: 10 x 10 x 45 mm
- Optional: cell holder 20 x 10 x 45 mm
- Power requirements: 110/220 VAC +/- 10 % - 50/60 Hz 30 watts
- Dimensions: height 12 cm; width 34 cm; depth 31 cm
- Net weight: 3.6 kg
- Weight with crate: 4.10 k
|
| |
|
|
METROLAB - 1600
DETECTOR
- Photo diode (320-1000 nm).
WAVELENGTH RANGE
- Automatic 9 positions filter wheel;
- Seven standard interference filters: 340, 405, 450, 505, 550, 620 and 750 nm.
- Three positions for optional filters.
TEMPERATURE CONTROL
- Peltier elements system.
- Temperature setting between 25ºC and 37ºC.
LANGUAGES
- English
- Spanish
- Chinese
- Portuguese
- Other languages on request.
POWER REQUIREMENTS
- 100-240 VAC 50/60 Hz;
- Ground connection
- 150 W
|
| |
|
|
METROLAB - 2300
Random Access Clinical Analyzer for Clinical Chemistry, Turbidimetry, Stats, Coagulation, Quality Control.
|
|
Samples: |
48 vials and/or primary tubes and/or
pediatric vials |
|
|
Sample volume programmable from 2 to
100 microliters |
|
Reagents: |
Tray admits 48 refrigerated reagents |
- Dispensing system
Active termostated dispensing system
Level detection by high frequency capacitive device.
Internal and external tip wash.
- Reaction tray
Includes 80 cuvettes for simultaneous incubation
Heating system with air circulation: Room temperature,
30 oC and 37 oC. Admits other calibration temperatures.
Available cuvettes: 6 and 10 mm light path.
- Optics
True double beam optical system.
Interference filters: 340, 380, 405, 450, 505, 550, 600 and 700 nm
Other wavelengths, on request.
Bandwidth: 10 nm
Photometric range: -0,1 to 3.6 absorbance units. (-0.1 to 5.0 with 6 mm cuvettes)
- Analysis modes
End point with sample or reagent blank
Standard or factor
Calibration curve with up to 10 standards.
Automatic curve fit and re-calculation.
Turbidimetry
Coagulation by turbidimetric method
Kinetic and fixed point (0th and 1st order rates)
Batchs, profiles, stats.
Pre-dilution of samples
Automatic dilution for abnormal values, depletion rate, prozone effect and linearity limits.
Storage and selection of samples by method, date, patient ID, etc.
Automatic backup
|
| |
|
|